Monitoring activity of amputee patients in rehabilitation using accelerometry: Validation study

Salih S, Peel NM, Burgess K

Australian & New Zealand Society for Geriatric Medicine 28 – 30 May 2014; Australas J Ageing 2014; 33 (S1) : 36 – 37

Aims: To validate the accuracy of tri axial accelerometer (ActivPAL) in measuring walking activity of amputee patients while using prosthetic limb. DESIGN: Observational study. STUDY SITE: Princess Alexandra Hospital, Geriatric and Rehabilitation Unit (GARU) in Brisbane. PARTCIPANTS: Twenty one subjects recruited from inpatient amputee ward and amputee outpatient clinic. Methods: Recruited patients were requested to wear accelerometers taped to the thigh of both the amputated and non amputated leg. A physiotherapist monitored and recorded the time of three activities patients were asked to perform: walking with prosthesis with or without walking aids for 5 minutes, self-propelling wheel chair for 3 minutes and pushed in wheel chair for 3 minutes. The device recorded amount of time in minutes spent walking. Data were downloaded from the device into spreadsheet. The data collected from observed routine activities was compared with accelerometer output to validate the accuracy of accelerometer data. Accuracy (%)= Measured Walking Time (s) Observed Walking Time (s) 1000. The standard error of the mean (SEM) was also calculated. Results: The mean age was 59.4 (SD 11.5), 15 patients were male. The majority (n = 17) had Trans Tibial Amputation (TTA) and 4 patients had Trans Femoral Amputation (TFA). The reason for amputation in 13 patients was vascular and traumatic in 4 patients. Twelve patients walked with nil aids, 5 patients with 2 SPS, 3 patients with SPS and 1 patient with 4 WW. The duration of time since last ?tting was variable, with 10 patients ?tted within one year, 8 patients between 1 5 years and for 3 patents between 5 to 36 years. The accuracy of the walking time extracted from accelerometer attached to the amputeelegwas97.6%, with mean 4.88, ranging from 3.91 to 5.08 minutes, with SEM of 0.01. The accuracy of the walking time extracted from accelerometer attached to non amputee leg was 99.8%, mean 4.99, ranging from 4.9 to 5.11 minutes, with SEM of 0.06. There were no walking signals detected when patients mobilised in a wheel chair. Conclusion: The use of an accelerometer in measuring walking time for an amputee using a prosthesis is valid. It has potentially important clinical application for prescribing prostheses.

NA

Salih S, Peel NM, Burgess K Monitoring activity of amputee patients in rehabilitation using accelerometry: Validation study Australian & New Zealand Society for Geriatric Medicine 28 – 30 May 2014; Australas J Ageing 2014; 33 (S1) : 36 – 37 2014. http://onlinelibrary.wiley.com/doi/10.1111/ajag.2014.33.issue-s1/issuetoc.