Comparison of 7 physical activity monitors: a validation study

Mazzà, C; Storm, F; Heller, B

ISPGR World Congress

BACKGROUND AND AIM: The number of steps walked per day, typically measured with activity monitors, can be considered as the most reliable daily living measurable input for non-exercise physical activity models. Accelerometry is the most commonly exploited technology in physical activity monitoring. A number of physical activity monitors are available on the market, for both clinical and generic purposes. The aim of this work was to compare 7 physical activity monitors, belonging to one of the two above categories, and assess their accuracy in step detection during three walking protocols at different speeds. Data were collected both in an indoor laboratory setting and outdoor daily life conditions. METHODS: Sixteen healthy adults (10 males, 6 females; age: 28.9 ± 2.7 years; height: 1.75 ± 0.09 m, mass: 72.0 ± 9.2 kg; BMI: 23.46 ± 2.34 kg/m2) were recruited to the study after having given informed consent. They were fitted with seven activity monitors comprising the DynaPort Movemonitor (McRoberts), UP (Jawbone), One (Fitbit), ActivPAL (PAL Technologies Ltd.), Tractivity (Kineteks Corp.), Nike+ Fuelband (Nike Inc.), Sensewear Armband Mini (Bodymedia), each of those was located following the manufacturers indications. Data was collected during three 11-minutes trials, in which the participants walked at their self-selected natural, slow, and fast speeds. Each trial entailed walking on a 20-m long indoor straight walkway, walking along a self-selected path in an indoor setting, and walking along an outdoor path. Two additional magneto-inertial sensors (OPAL, ADPM Inc) were located on each of the participants’ ankles and provided a definitive reference for the step count, with heelstrikes detected from the angular velocity around the mediolateral axis using a previously validated algorithm [1]. Total count of left and right steps (N) was then computed. Differences in group estimates between sensor outcomes were tested. Bland-Altman plots were used to test agreement between variables and to evaluate bias between scores. RESULTS: In all the speed conditions, N was underestimated by the DynaPort, Fitbit, ActivPAL, Fuelband and Armband (p<0.05) and overestimated by the Tractivity. All sensors performed best at the fastest walking speed. The lowest absolute percentage error was found for the DynaPort (<2.0%, Bland-Altman plot shown in figure 1), followed by ActivPAL and Fitbit One (errors <2.6% and 3.2%, respectively). CONCLUSIONS: This study provides a robust comparison of activity monitors to support the choice of a device for step recognition in a healthy, young adult population. DynaPort and ActivPAL were the best performing devices among those for clinical applications examined in this study. Fitbit One is a lower-cost alternative with only slightly inferior performance. ACKNOWLEDGMENTS: This study was funded by the EC contract FP7-ICT-2011-9, No.600803 (MISSION-T2D), and by the Sheffield National Centre for Sport and Exercise Medicine.

NA

Mazzà, C; Storm, F; Heller, B Comparison of 7 physical activity monitors: a validation study ISPGR World Congress 2014. https://d3lut3gzcpx87s.cloudfront.net/down/eJwFwQEKgDAIBdAbaWE66jaf1ihqbaQQdPre2yP6wuzHdXrgCXIhVHztxuu0tsoqyFsxUZF5SGpI46RioJ7LD6rNFBQ=/Full Poster Abstract List.pdf.